System Rainbow
System Rainbow

Software control

Sample Drawer &
Autocalibration Part

Supercontinuum Laser Output

AUTOCALIBRATION

Real-time settings eliminate the need for complex recalibrations or costly technical services. 

MORE ACCURACY AND SPEED

Optimal results with Rainbow’s max performance—no speed limitations, only unmatched efficiency.

FRIENDLY USER INTERFACE

Unlock versatility as Rainbow interface monitors, tracks, and exports data easily. 

New in-line color Standard

System Rainbow

TECHNICAL SPECIFICATIONS

Measurement Zone Z1 Z2 Z3
Geometry* -14º / 65º 0º / 45º 14º / 23º
Measurement distance (machine aperture to sample) > 70 mm
Light source Supercontinuum white light laser
Spectral measurement range 400 nm - 750 nm
Measurement area** 60 mm 60 mm 60 mm
Total measurement zone 450 mm
Measurement time < 100 ms < 100 ms < 100 ms
Minimum measurement interval 2 s between zones
Distance variation with error ΔL*, Δa, Δb < 0.4 ± 2 mm ± 2 mm ± 2 mm
Repeatibility (standard deviation for 10
repeated measurements of a white sample)***
ΔL*, Δa, Δb, ≤ 0.1 ΔL*, Δa, Δb, ≤ 0.1 ΔL*, Δa, Δb, ≤ 0.1
Reproducibility (standard deviation for 10
different measurements of a white sample over 1 hour)
ΔL*, Δa, Δb, ≤ 0.2 ΔL*, Δa, Δb, ≤ 0.2 ΔL*, Δa, Δb, ≤ 0.2
Calibration (traceable to NIST/PTB with
certificate for the 250nm – 2450nm range)
Automatic calibration in each measurement zone
Environment light influence NO

*Incidence referred to the normal/detected to the specular.

**Customizable upon request from 10 mm up to 80 mm.

***By taking continuous measurements over 24 hours, the repeatibility of ∆L*, ∆a, and ∆b are ≤ 0.2.

MECHANICAL SPECIFICATIONS

Environment temperature 5 - 40 °C
Machine size 976 mm x 2783 mm x 1020 mm
Machine weight 450 kg
User Interface Specific HMI for coil coating industry (Color display 17")
I/O Interface Ethernet, RS232, RS485, USB
Input voltage 230 V AC
Max power consumption 3.6 kW